The fan Respironics V60 combines experience in the ventilation Respironics Philips approach to health care advanced and easy. The result is the gold standard of noninvasive ventilation with a safety net and an interactive display that helps simplify the management of patients.
Alarms
Alarm Adjustable range
Freq. high (high respiratory rate alarm) 5 - 90 RPM
Freq. low (low respiratory rate alarm) 1-89 RPM
High VT (high tidal volume alarm) 200 - 2,500 ml
Low VT (low tidal volume alarm) Off - 1,500 ml
PIA (high inspiratory pressure alarm) 5 - 50 cmH2O
GDP (low inspiratory pressure alarm) Off, 1 - 40 cmH2O
Low VE (minute ventilation alarm low) Off, 0.1 to 99 l / min
GDP T (delay time low inspiratory pressure) 5-60 s |
Parameters monitored
Patient Data Window
Phase Indicator / on spontaneous breathing., Temp., Exhale
PIP 0-50 cmH2O
Overall leakage / patient from 0 to 200 l / min BTPS
Activ. pac. 0 - 100%
Respiratory rate from 0 to 90 RPM
Ti / Ttot 0-91%
Minute volume from 0 to 99.0 l / min BTPS
`Tidal Volume 0-3000 ml BTPS
Waveform Window
Waveform Press 0 - 50 cmH2O
Flow waveform from -240 to 240 l / min BTPS
Waveform Volume 0-3000 ml BTPS |
Other adjustments
Alarm Volume 1 to 10 (relative scale)
Brightness 1 to 5 (relative scale)
Port Selection disposable exhalation exhalation port (PED)
Whisper Swivel
Plateaux exhalation valve (VEP)
Other
None (no exhalation port circuit Online)
Interface Selection ET / tracheal, 1, 2, 3, Other
Screen Lock Off, On |
Environmental
Temperature
Operating condition +5 to +40 º C
Terms of -20 to +50 º C
Humidity
Operating conditions 15 to 95% (no condensation)
Storage conditions 10 to 95% (no condensation)
Barometric Pressure
Operating and storage from 79.9 to 101.1 kPa
(600-765 mmHg)
Altitude
Operating and storage -51 to 1951 m (-167 to 6400 feet)
relative to sea level |
Electrical
External
Voltage AC 100 to 240 VAC
AC Frequency 50 to 60 Hz
300 VA AC Power
Battery (optional)
14.4 V nominal value
Capacity 11.5 Ah
+16.9 V load voltage maximum
Duration 6 hours under normal |
General
Demand Flow range from -240 to 240 l / min BTPS
Inlet pressure range of oxygen
40 to 87 psig (276 to 600 kPa)
Input Connectors DISS oxygen (male or female), NIST (male), SIS (male)
Dynamic pressure regulation ± (2 cmH2O + 4% of target)
Start Ready for airing 9 seconds
after power
Acoustic Noise <45 dBA at 1 m |
Physical
Weight 10.9 kg (24 lb)
with optional battery
10.0 kg (22 lb)
without optional battery
Dimensions 33.7 cm (13.3 inches) high
39.4 cm (15.5 inches) wide
42.9 cm (16.5 in.) deep
|