The BiPAP noninvasive ventilator Focus is a basic management specifically designed for bilevel serves institutional settings for ventilatory support in patientswith stable and non-acute insufficiency or respiratory failure. The BiPAP Focussystem, compact and lightweight with features that make administration of noninvasive ventilation easy and effective.
Viewing patient parameters
This valuable feature gives the physician the convenience of viewing all patient parameters at once. |
Bar graph of the pressure
Quickly and clearly indicates the patient's breathing and pressure ranges of the patient. |
Alarm easy to use
Pressure alarms are automatically set at 5 cm H2O above the IPAP and EPAP below the set, while minimizing nuisance alarms and increasing patient safety. |
Backup Battery
A comprehensive backup battery keeps the patient's ventilation in case power is interrupted. This security feature also provides a practical method for in-hospital transport of patients without interrupting ventilation. |
Continuum of care
The BiPAP Focus is part of our solutions to the continuum of care (Continuum of Care) for that help, it monitors and controls to patients with respiratory disorders. |
Pressure: CPAP 4 to 20 cm H2O (4 to 20 hPa)
Resolution: 1 cm H2O (1 hPa)
Dynamic Accuracy: ± 5 cm H2O (± 5 hPa)
EPAP 4 to 20 cm H2O (4 to 30 hPa)
Resolution: 1 cm H2O (1 hPa)
Dynamic Accuracy: ± 5 cm H2O (± 5 hPa)
IPAP 4 at 30 cm H2O (4 to 20 hPa)
Resolution: 1 cm H2O (1 hPa)
Dynamic Accuracy: ± 5 cm H2O (± 5 hPa) Rise Time: Set 1 to 6 (where 1 = 0.1 seconds and 6 = 0.6 seconds)
Resolution: 1
Accuracy: ± 10% of the configuration (R = 50, C = 20)
Frequency: 1 to 30 BPM
Resolution: 1 BPM
Accuracy: ± 1 BPM or ± 10% of setting, whichever is greater over a period of 4 minutes.
Configurations of Time and frequency-I can not allow the time-I exceeds the expiratory time.
Time-I: 0.5 to 3 seconds
Resolution: 0.1 seconds
Accuracy: ± 0.1 +10% of the second set
Configurations of Time and frequency-I can not allow the time-I exceeds the expiratory time.
Ramp time: 0 to 45 minutes
Resolution: 5 minutes
Accuracy: ± 10% of the set
Start Ramp: 4 cm H2O to EPAP (4 hPa EPAP)
Resolution: 1 cm H2O (1 hPa)
Patient parameters
Patient circuit pressure: Range: 0 to 35 cm H2O (0 to 35 hPa) (continuous display of the bar graph)
Frequency Range: 0 to 99 BPM
Vt Estimated range: 0 to 4,000 ml
Estimated MV: Range: 0 to 99 l / min
Patient Leakage: Range: 0 to 200 l / min
Frequency of apnea: Range: 0 to 99 hours
Alarms
High system pressure, 5 cm H2O (5 hPa) on the IPAP pressure target (auto)
Low system pressure: 5 cm H2O (5 hPa) under the EPAP pressure target (auto)
Apnea Interval: 20, 40, 60 seconds or off
Frequency of apnea: Range: 5, 10, 20 apneas per hour or off
Call the nursing staff: Suitable for systems normally open or normally closed
Physical Characteristics
Dimensions: 140 high x 290 wide x 355 deep (mm)
Weight: 4.5 kg
Feeding
Power input: 100 V AC and 50/60 V CA/230 Hz
Battery Standby: 3.8 amp-hours (NiMH)
Fuse: Fuse not replaceable
Call the nursing staff: Rated current: 0.280 A
Rated voltage: 250 V
|